GDOT-80A Qalabka Tijaabada Saliida Insulation-ka waa la cusboonaysiiyay1105
●Fadlan si taxadar leh u akhri buug-gacmeedka hawlgalka ka hor inta aanad shaqayn.
●Fadlan hubi in tijaabiyaha uu si fiican ugu xiran yahay dhulka ka hor inta aadan tijaabin.
●Waa mamnuuc in la dhaqaajiyo ama kor loo qaado daboolka tijaabada inta lagu jiro habka tijaabada si looga fogaado dhaawaca koronta sare.Korontadu waa in ay dansataa ka hor inta aan la beddelin saamiga saamiga.
●Si taxadar leh ula tacaal markaad iska bixinayso ama xidhayso daboolka tijaabada tamarta sare!
●Haddii tijaabiyaha uu u shaqeeyo si aan caadi ahayn ka dib marka saliidda dahaadhka ahi ay jabto, fadlan dami tijaabiyaha ilaa 10 ilbiriqsi, ka dibna dib u bilow.
●Kadib marka warqadda daabacadu dhammaato, fadlan tixraac qaybta sharraxaadda daabacaha (ama lifaaqa gacanta) si aad u beddesho warqadda daabacaadda si aad uga fogaato dhaawaca madaxa daabacaha.
●Tijaabiyaha ka fogee qoyaanka, boodhka iyo walxaha kale ee daxalka, kana fogee meelaha heerkulku sarreeyo.
●Si taxadar leh ula dhaqan gaadiidka.Hoos ha u dhigin.
●Buuggu waxa uu dib u eegi karaa si waafaqsan ogaysiis dheeraad ah oo hore.Haddii wax su'aal ah, fadlan nala soo xiriir.
Muddada dammaanadda ee taxanahan waa hal sano laga bilaabo taariikhda shixnadda.Fadlan tixraac qaansheegtaada ama dukumeentiyada rarida si aad u go'aamiso taariikhaha dammaanadda ee habboon.Shirkadda HVHIPOT waxay dammaanad qaadaysaa iibsadaha asalka ah in alaabtani ay ka xoroobi doonto cilladaha alaabta iyo shaqada ee isticmaalka caadiga ah.Inta lagu jiro xilliga dammaanadda, bixi in cilladaha noocaas ah aysan go'aamin HVHIPOT inay ka dhasheen xadgudub, si xun u isticmaalka, beddelka, rakibaadda aan habboonayn, dayacaad ama xaalad deegaan oo xun, HVHIPOT waxay ku xaddidan tahay oo keliya hagaajinta ama beddelka qalabkan inta lagu jiro muddada dammaanadda.
Qalabka GDOT-80C | 1 pc |
Koob saliid (250ml) | 1 pcs |
Xadhiga korantada | 1 pc |
fiyuuska kaydka ah | 2 pcs |
Usha walaaqaya | 2 pcs |
Qiyaasta caadiga ah (25mm) | 1 pc |
Warqad daabac | 2 duub |
Tweezer | 1 pc |
Tilmaamaha isticmaalaha | 1 pc |
Warbixinta tijaabada warshadda | 1 pc |
HV Hipot Electric Co., Ltd. ayaa si adag oo taxadar leh u akhriyay buug-gacmeedka, laakiin ma dammaanad qaadi karno in aanay jirin wax khalad ah iyo khaladaad gebi ahaanba ku jira buug-gacmeedka.
HV Hipot Electric Co., Ltd. waxaa ka go'an in ay si joogto ah u horumariso hawlaha alaabta, iyo hagaajinta tayada adeegga, sidaas darteed shirkadu waxay weli xaq u leedahay inay beddesho wax kasta oo badeecooyin ah iyo barnaamijyada software ee lagu qeexay buug-gacmeedkan iyo sidoo kale waxa ku jira buug-gacmeedkan iyada oo aan horay loo arag. ogeysiis.
Xakamaynta gudaha ee qalab koronto oo badan oo ku jira nidaamyada korantada, nidaamyada tareenada, dhirta batroolka ee waaweyn iyo shirkadaha, ayaa inta badan la qaatay nooca dahaarka saliidda ka buuxo, sidaas darteed, tijaabinta tamarta dielectric saliid insulating waa mid caadi ah oo lagama maarmaan ah.Si loo daboolo baahiyaha suuqa, waxaanu soo saarnay oo aanu soo saarnay taxane ah oo ah tijaabiyeyaal awood-dielectric saliid ah oo dahaaran iyadoo loo eegayo heerka qaranka GB/T507-2002, heerka warshadaha DL429.9-91 iyo heerkii ugu dambeeyay ee warshadaha tamarta korontada DL/T846.7 -2004 nafteena.Qalabkan, adoo isticmaalaya microcomputer hal-chip ah oo udub dhexaad u ah, wuxuu xaqiiqsaday hawlgal toos ah, cabbir sax ah oo sarreeya, si weyn u hagaajinaya waxtarka shaqada iyo yaraynta xoogga shaqada ee shaqaalaha.Waxaa intaa dheer, waa mid yar oo cabbir ah oo ku habboon in la qaado.
●Iyada oo microprocessor ah, si toos ah u buuxi tijaabada tamarta adkeysiga ee wareegga saliidda oo leh 0 ~ 80KV (oo ay ku jiraan kor u qaadista, ilaalinta, isku dhafka, istaagida, xisaabinta, daabacaadda iyo hawlgallada kale).
●Shaashad weyn oo LCD ah.
●Hawlgal fudud.Mashiinku wuxuu si otomaatig ah u dhammaystiri doonaa tijaabada korantada ee hal koob oo saliid muunad ah ka dib marka uu dejiyo hawlwadeenku.Qiimaha korantada ee 1 ~ 6 jeer iyo wakhtiyada wareegyada ayaa si toos ah loo kaydin doonaa.Imtixaanka ka dib, daabacaha kulaylka ayaa daabici doona qiimo jabka kasta iyo celceliska qiimaha.
●Ilaalinta awoodda hoos u dhigista.Waa awoodbadbaadi 100 natiijooyin tijaabo ah oo muuji heerkulka deegaanka iyo qoyaanka hadda.
●Qabo kombuyuutar hal-chip ah si aad kor ugu qaaddo xawaaraha joogtada ah.Inta jeer ee danabku waa sax at 50HZ, hubi in habka oo dhan ay fududahay in la xakameeyo.
●Iyada oo korantada xad dhaafka ah, hadda xad dhaafka ah iyo ilaalinta xaddidan si loo xaqiijiyo badbaadada hawlwadeennada.
●Iyadoo shaqada muujinaysa heerkulka la qiyaasay iyo saacadda nidaamka.
●Kumbiyuutarka kula xidhiidh interface-ka caadiga ah ee RS232.
Korontada | AC220V± 10%, 50Hz |
Korontada wax soo saarka | 0~80kV(La dooran karo) |
Awood | 1.5kVA |
Awood | 200W |
Xawaaraha korontadu | 2.0~3.5kV/s (la hagaajin karo) |
Voltagecabbiriddasaxnaanta | ± 3% |
Waveform qalloocan | ≤3% |
Kordhinta muddada | 5 daqiiqo (La hagaajin karo) |
Waqtiga taagan | 15 daqiiqo (La hagaajin karo) |
Waqtiyada kobcinta | 1~6 (La dooran karo) |
Hawlgalkadeegaanka | Temperature: 0℃-45°C |
Hxudunta:Max.qoyaan qaraabo75% | |
Cabirka | 465x385x425mm |
① Daabacaha kuleylka - daabacaadda natiijooyinka imtixaanka;
② LCD--muujinaya liiska, degdega ah iyo natiijooyinka tijaabada;
③ Furayaasha shaqada:
Riix "◄" furaha si aad u kordhiso qiimaha dejinta;
Taabo furaha "►" si aad u yaraato qiimaha dejinta;
Dooro - si aad u doorato hawlaha ( shayga la doortay ayaa yaala bandhiga kaydka);
Xaqiiji - fulinta hawlaha;
Dib-u-soo-baxa ka-baxa interface-ka hawlgalka;
④ Korontada iyo tilmaame
1. Diyaarinta imtixaanka ka hor
1.1 Ku xidh terminalka dhulka hoostiisa (dhanka midig ee qalabka) si adag siligga dhulka ka hor inta aanad isticmaalin qalabka.
1.2 Soo saar muunadda saliidda si waafaqsan heerka ku habboon.Ku hagaaji fogaanta korantada gudaha koobka saliidda sida qiyaasta caadiga ah.Ku nadiifi koobka si waafaqsan shuruudaha khuseeya.Ku shub saamiga saliidda koobka oo xidh furka.
1.3 Ku beddelashada tamarta AC220V ka dib marka la xaqiijiyo walxaha kor ku xusan, diyaar u ah imtixaanka.
2. Tijaabinta
2.1 Riix furaha korantada ka dibna gal interface-ka soo socda:
2.2 Habaynta Halbeegga Nidaamka
Riix "Confirm" furaha oo geli interface soo socda:
Dejinta kobcinta: isticmaalayaashu waxay dooran karaan si waafaqsan baahida dhabta ah.
Riix "Back" furaha si aad uga baxdo interface this ka dib marka dejinta la sameeyo.
2.3 Tijaabinta
Riix furaha "Select" si aad u dooratid "Start Test" menu oo taabo furaha "Confirm" si aad u gashid interface soo socda:
Si aad u sii wadato imtixaanka soo socda isla marka tijaabada ugu horeysa dhamaato ilaa inta jeer ee kor u qaadida la dhammeeyo.Ugu dambeyntii, natiijadu waa la muujiyay oo loo daabacay sida soo socota:
2.4 Daawashada iyo daabacaadda xogta:
Riix furaha "Dooro" si aad u dooratid "Daabacaadda Xogta iyo Daabacaadda" menu oo taabo furaha "Confirm" si aad u gasho interface-ka soo socda:
Dooro "Bogga Up" ama "Bogga Down" oo dooro diiwaannada la daabacayo oo dooro "Daabac".
●Xulashada muunadda saliidda iyo meelaynta fogaanta korantada waa inay buuxisaa heerarka qaranka iyo warshadaha ee khuseeya.
●Hawl-wadeenada ama shaqaalaha kale waxa si adag looga mamnuucay inay taabtaan qolofka ka dib marka korontada la shido si looga fogaado shil.
●Korontada waa in isla markiiba la gooyaa haddii wax aan caadi ahayn la helo inta lagu jiro hawlgalka.
●Qalabkani waa inaan lagu soo bandhigin deegaanka qoyan.
●Koobka saliidda iyo korantada ka dhig nadiif.Buuxikoob leh saliid transformer cusub si ay u ilaaliso marka ay shaqaynayso.Fiiri fogaanta korantada oo hubi cidhiidhiga u dhexeeya caarada korantada iyo xadhkaha korantada ka hor inta aan koobka mar kale la isticmaalin.
1. Habka Nadiifinta Koobka Saliidda
1.1 Ku masax dusha elektiroonigga ah iyo baararka mar labaad iyo mar kale maro xariir ah oo nadiif ah.
1.2 Ku hagaaji masaafada korantada iyo cabirka caadiga ah
1.3 Isticmaal ether batroolka (waxyaabaha kale ee dabiiciga ah waa mamnuuc) si aad u nadiifiso saddex jeer.Waqti kasta waa inuu raacaa hababka soo socda:
① Ku shub weelka batroolka koobka saliidda ilaa koobku ka buuxo 1/4~1/3.
② Ku dabool cidhifka koobka gabal quraarad ah oo ay nadiifisay eter batroolka.Ku rux koobka si siman hal daqiiqo xoog gaar ah.
③ Ku shub weelka batroolka oo ku qalaji koobka afuufiyaha 2 ~ 3 daqiiqo.
1.4 Isticmaal saamiga saliidda si aad u nadiifiso koobka 1 ~ 3 jeer.
① Ku shub weelka batroolka koobka saliidda ilaa koobku ka buuxo 1/4~1/3.
② Ku dabool cidhifka koobka gabal quraarad ah oo ay nadiifisay eter batroolka.Ku rux koobka si siman hal daqiiqo xoog gaar ah.
③ Ku shub saamiga saliidda bidix ka dibna baaritaanka ayaa bilaabmaya.
2. Habka Nadiifinta Usha Kicinta
2.1 Mar labaad iyo mar kale ku tirtir usha wax kicinaysa maro nadiif ah oo nadiif ah ilaa qaybo yaryar laga waayo dushooda.Waa mamnuuc in gacmaha lagu taabto sagxadaha.
2.2 Isticmaal xoog si aad usha u qabato;ku rid eter batroolka oo ku dhaq.
2.3 Adeegso xoog si aad ugu xidho usha oo aad ku qalajiso afuufiye.
2.4 Isticmaal xoog si aad usha u qabato;ku rid saamiga saliidda oo ku dhaq.
3. Kaydinta Koobka Saliidda
Habka 1 Ku buuxi koobka saliid dahaadh wanaagsan ka dib marka baaritaanka la dhammeeyo oo dhig meel deggan.
Habka 2 Nadiifi oo qalaji koobka hoos hababka kor ku xusan ka dibna ku rid vacuum qalajiyaha.
Fiiro gaar ah: Koobka saliidda iyo usha wax lagu karsado waa in lagu nadiifiyaa hababka kor ku xusan ka dib baaritaanka ugu horreeya oo lagu tijaabiyo saliid xun.
4. Nadiifinta Cilada Guud
4.1 Korontadu way dansan tahay, wax shaashad ah oo ka muuqdana ma jiraan
① Hubi in bareega korontada la geliyey si adag iyo in kale.
② Hubi in fiyuuska ku dhex jira korantada in uu xaalad fiican ku jiro iyo in kale.
③ Hubi korontada godka.
4.2 Koob saliid oo aan ifafaale burbur lahayn
① Hubi gelinta xidhiyaasha sabuuradda wareegga.
② Hubi xidhiidhka furaha korantada sare ee daboolka kiiska.
③ Hubi firfircoonida xiriirada tamarta sare leh.
④ Hubi jebinta khadka korantada sare.
4.3 Kala duwanaanshaha bandhiggu kuma filna
Hagaji potentiometer-ka sabuuradda wareegga.
4.4 Burburka daabacaadda
① Hubi furka korantada daabacaha.
② Hubi xidhitaanka khadka xogta daabacaha.